RCF to RPM Calculator
Convert between relative centrifugal force (× g) and revolutions per minute for your rotor's radius, either way, with ten rotors from Eppendorf and Beckman Coulter ready to choose. Move a protocol written for another rotor to yours: the speed that gives the same force, and, from the two k-factors, the time that pellets the same particles.
Calculator
RCF and rpm
The relative centrifugal force, in × g, from the speed in rpm and the rotor's radius.
Choose one to fill in its radius and see its top speed, or type your rotor's radius below.
From the axis to the bottom of the tube, as the rotor's manual gives it.
Revolutions per minute, as the centrifuge is set.
Enter the speed and radius, or choose a rotor: the force in × g comes here.
Worked example
13,000 rpm in an Eppendorf 5424 R, rotor FA-45-24-11
- 01r max is 8.4 cm, from the rotor's manual
- 02ω = 2π × 13,000 ÷ 60 = 1,361.36 rad/s
- 03a = 0.084 m × 1,361.36² = 155,677 m/s²
- 04RCF = 155,677 ÷ 9.80665 = 15,875 × g
- 05In one step: 1.118244 × 10−5 × 8.4 × 13,000² = 15,875 × g
15,875 × g. The same 13,000 rpm in a MiniSpin, whose radius is 6 cm, would give only 11,339 × g.
How it works
How to convert RPM to RCF
A spinning rotor pushes the sample outwards with an acceleration of r ω², where ω is the angular speed in radians a second. Relative centrifugal force is that acceleration in multiples of the earth's own, 9.80665 m/s², so the figure has no units: 15,875 × g means the bottom of the tube is pulled 15,875 times as hard as it is by gravity. Putting ω = 2π × rpm ÷ 60 and the radius in centimetres into r ω² ÷ g gives the one line every manual prints, RCF = 1.118 × 10−5 × r × rpm².
Force goes with the square of the speed and only with the first power of the radius, so halving the speed quarters the force, while the radius changes it in proportion. That is why the same number on two centrifuges is not the same spin: the radius is part of the answer, and a protocol should give × g.
Going the other way is the same line rearranged: rpm = √(RCF ÷ (1.118 × 10−5 × r)). Because most centrifuges are set in steps of 100 rpm, the calculator shows the nearest hundred either side and what each gives, so you can pick the one you can actually dial in.
Pelleting depends on how far a particle has to travel as well as how hard it is pushed, which is what the k-factor, or clearing factor, measures: k = 2.533 × 1011 × ln(r max ÷ r min) ÷ rpm², the hours it takes to pellet a particle of one svedberg. A smaller k clears faster. To run a protocol on a different rotor, match the force and scale the time by the two k-factors, t₂ = t₁ × k₂ ÷ k₁.
Questions
Frequently asked
How do you convert RPM to RCF?
Multiply the radius in centimetres by the speed squared and by 1.118 × 10−5: RCF = 1.118 × 10−5 × r × rpm². At 13,000 rpm with a radius of 8.4 cm that is 15,875 × g. Use the maximum radius of the rotor, from the axis to the bottom of the tube, which is in the rotor's manual.
How do you convert RCF to RPM?
Take the square root: rpm = √(RCF ÷ (1.118 × 10−5 × r)), with the radius in cm. For 12,000 × g at a radius of 7.7 cm, rpm = √(12,000 ÷ (1.118 × 10−5 × 7.7)) = 11,805 rpm, so you would set 11,800 or 11,900.
What is relative centrifugal force?
The force a spinning rotor puts on the sample, written as a multiple of the earth's gravity: 1,000 × g means a thousand times its own weight. It is r ω² ÷ 9.80665 m/s², so it depends on both how fast the rotor turns and how far the sample sits from the axis. It has no units, which is why it is the same figure on any centrifuge.
What is the difference between RCF and RPM?
RPM is how fast the rotor turns; RCF is how hard the sample is pushed, and it depends on the radius as well as the speed. The same 10,000 rpm is 6,709 × g in a 6 cm microcentrifuge rotor and 18,563 × g in a 16.6 cm floor rotor. RCF, g-force and × g all name the same figure, which is why protocols give it: anyone can work out the speed their own rotor needs.
Which radius should I use, r max or r min?
r max, from the axis to the bottom of the tube, unless the protocol says otherwise: it is where the pellet forms and what makers quote their rotors' maximum force at. r min is the top of the sample and r av halfway between; they matter for gradients, and r min and r max together give the k-factor.
What is a rotor's k-factor?
Its clearing factor: the hours it would take to pellet a particle of one svedberg at a given speed, k = 2.533 × 1011 × ln(r max ÷ r min) ÷ rpm². The time for a real particle is k ÷ S, and to move a run to another rotor at the same force, t₂ = t₁ × k₂ ÷ k₁. A smaller k clears faster.
Reference
RPM to × g at common radii
Relative centrifugal force at each speed and radius, 1.118 × 10−5 × r × rpm2. Microcentrifuge rotors are about 6 to 8.5 cm, benchtop swinging buckets 15 to 20. Find your rotor's own radius in its manual, or choose it in the calculator.
| rpm | 6 cm | 8 cm | 15 cm |
|---|---|---|---|
| 500 | 17 | 22 | 42 |
| 1,000 | 67 | 89 | 168 |
| 1,500 | 151 | 201 | 377 |
| 2,000 | 268 | 358 | 671 |
| 2,500 | 419 | 559 | 1,048 |
| 3,000 | 604 | 805 | 1,510 |
| 4,000 | 1,074 | 1,431 | 2,684 |
| 5,000 | 1,677 | 2,236 | 4,193 |
| 6,000 | 2,415 | 3,221 | 6,039 |
| 8,000 | 4,294 | 5,725 | 10,735 |
| 10,000 | 6,709 | 8,946 | 16,774 |
| 12,000 | 9,662 | 12,882 | 24,154 |
| 13,000 | 11,339 | 15,119 | 28,347 |
| 14,000 | 13,151 | 17,534 | 32,876 |
| 15,000 | 15,096 | 20,128 | 37,741 |
In × g, at the radius given, worked out with the exact constant, 1.118244 × 10−5. A manual's own table may differ in the fourth digit, where it rounds the constant to 1.118 × 10−5.
Rotors
The rotors in the calculator
Ten rotors whose radius, top speed and force their makers publish. Choosing one in the calculator fills in its radii and warns when a speed passes its maximum. Every figure below is the maker's; the force each one gives at its top speed agrees with the formula to within half a per cent.
| Rotor | r max | Top speed |
|---|---|---|
| MiniSpin · F-45-12-11Eppendorf · Fixed angle | 6 cm | 13,40012,100 × g |
| MiniSpin plus · F-45-12-11Fixed angle | 6 cm | 14,50014,100 × g |
| 5418 R · FA-45-18-11Fixed angle | 7.7 cm | 14,00016,873 × g |
| 5424 R · FA-45-24-11Fixed angle | 8.4 cm | 15,00021,130 × g |
| SX4750Beckman Coulter · Swinging bucket | 20.78 cm | 4,7505,250 × g |
| JLA-10.500Fixed angle · k 2,850 | 16.6 cm | 10,00018,600 × g |
| JA-25.50Fixed angle · k 418 | 10.8 cm | 25,00075,600 × g |
| Type 45 TiFixed angle · k 133 | 10.38 cm | 45,000235,000 × g |
| SW 32 TiSwinging bucket · k 204 | 15.25 cm | 32,000175,000 × g |
| SW 41 TiSwinging bucket · k 124 | 15.31 cm | 41,000288,000 × g |
- Eppendorf: MiniSpin · F-45-12-11, 5418 R · FA-45-18-11, 5424 R · FA-45-24-11
- Beckman Coulter: SX4750, JLA-10.500, JA-25.50, Type 45 Ti, SW 32 Ti, SW 41 Ti
Radii are to the bottom of the tube, without an adapter. k-factors are each rotor's at its top speed; at any other speed, k goes up with the square of the ratio of the speeds. Always follow your own rotor's manual, which is the only place its derating and tube limits are written.
More Calculators
Each one shows its formula and a worked example, so you can check the answer and learn the method.
Hemocytometer Calculator
Cells per mL, viability with trypan blue, total cells and the volume to seed, from a hemocytometer count.
Open calculatorLigation Calculator
How much insert to add to your vector for a molar ratio, with the T4 DNA Ligase or assembly reaction to pipette, and DNA between ng, pmol and copies.
Open calculatorSerial Dilution Calculator
The volume to transfer, the diluent for each tube and the concentration in every tube of a dilution series.
Open calculatorPractice the theory with free quizzes in Cell Biology, Molecular Biology, Microbiology, Biochemistry, Biotechnology and Immunology.